Promega Corporation

Modulus® Microplate Multimode Reader Software and Firmware

Need Assistance? Chat

  • Share
  • Print
  • Email

Download the latest Modulus® Microplate Instrument Firmware (version 01.72) and Software (version 1.2.0)

Important: This software and firmware update should only be performed on instruments with firmware version 01.61 or 01.62.

If the firmware version on the instrument is not one of those listed above, do not perform the update and contact Promega Technical Services support for assistance.

To perform the update, follow these steps:

  1. Copy the executable file (*.exe) provided to the root of the USB stick and insert into the USB drive on the instrument.
  2. Press the Ok button when asked to update software.
  3. Remove executable file from the USB stick after process is complete.
  4. Copy the firmware files (af.ldf, ff.ldf and fc.ldf) provided to the root of the USB stick and insert into the USB drive on the instrument.
  5. Press the Ok button when asked to update firmware.
  6. Remove the firmware files from the USB stick after process is complete.
  • Warning
  • It appears that you have Javascript disabled. Our website requires Javascript to function correctly. Please enable Javascript to experience our website.